QC Troubleshooting Guide

A systematic approach to identifying, investigating, and resolving QC failures in the clinical laboratory

Interactive Decision Flowchart

Follow this step-by-step guide to diagnose your QC issue. Click on each node to see detailed guidance.

START Which Rule Violated? 1-2s or 1-3s (Random Error) 2-2s or R-4s (Systematic Error) 4-1s or 10x (Trend/Extreme) Single or Multiple Levels? Single Level Likely random error All Levels Systematic error Investigate Root Cause See Investigation Checklist Below Document & Take Corrective Action

How to Use This Flowchart:

Click on any decision node (yellow diamonds) to explore that branch and see detailed guidance specific to your QC issue.

The flowchart guides you from rule violation detection → pattern recognition → root cause investigation → corrective action.

Error Classification Guide

Understanding the type of error is the first step toward resolution.

σ

Random Errors

Definition: Unpredictable variations that increase imprecision. Results scatter above and below the mean.

Characteristics:

  • Single level typically affected
  • Rules: 1-2s, 1-3s, R-4s violations
  • Random scatter pattern
  • Both directions of deviation

Common Causes:

  • Pipetting errors or technique variation
  • Air bubbles in samples
  • Temperature fluctuations
  • Clots in control material
  • Inadequate reagent mixing
  • Optical issues (cuvette contamination)

Investigation Focus:

  • Review operator technique
  • Check pipette calibration
  • Inspect control material condition
  • Verify temperature stability

Systematic Errors

Definition: Consistent deviation in one direction affecting accuracy. Results shift or trend away from target.

Characteristics:

  • Multiple/all levels affected equally
  • Rules: 2-2s, 4-1s, 10x violations
  • Consistent direction of deviation
  • Persists across multiple runs

Common Causes:

  • Calibration drift
  • Reagent lot change or deterioration
  • Environmental changes (temp, humidity)
  • Instrument maintenance effects
  • Decay of control material

Investigation Focus:

  • Check calibration status
  • Review reagent lot changes
  • Assess control material age
  • Monitor environmental conditions
!

Pre-analytical Errors (Mimicking QC Issues)

Definition: Sample problems that appear as QC failures but are not analytical issues.

How to Distinguish:

  • QC out of control but patient results unaffected
  • Only specific analytes affected
  • Control material age/expiration approaching
  • Recent freeze-thaw cycles of controls
  • Improper reconstitution of lyophilized controls

Action Items:

  • Replace control material with new lot
  • Verify proper reconstitution procedure
  • Check storage conditions
  • If QC resolves, document as control-related
  • If QC persists, escalate to systematic error investigation

Shift vs Trend: Visual Reference

SHIFT

Time → Target

Sudden, sustained jump. All points consistently above or below target. Often follows calibration change or reagent lot switch.

TREND

Time → Target

Gradual progression. Consistent drift direction over multiple runs. Indicates decay or gradual calibration loss.

Root Cause Analysis Tool

Use this interactive tool to get targeted suggestions based on your specific QC pattern.

QC Failure Investigation Checklist

Follow this step-by-step process to systematically identify the root cause.

Common QC Problems & Solutions Database

Search or browse 30+ common QC scenarios to find solutions quickly.

Observed Pattern Rule Violated Most Likely Cause Solution
Gradual upward trend over 10+ days 4-1s, 10x Calibration drift or reagent deterioration Recalibrate instrument. If drift persists, replace reagent lot. Implement more frequent calibration.
Sudden jump up after reagent lot change 2-2s or 1-3s Lot-to-lot reagent variability Run quality verification curve with new lot. Adjust calibration if needed. Document lot transition in QC record.
Random scatter, R-4s violation, single level R-4s Random pipetting error or control instability Review operator technique. Verify control reconstitution. Re-run QC with fresh control. Train operator if needed.
Both high and low controls high after calibration 2-2s (both levels) Calibration overcompensation or reagent issue Review calibration procedure. Verify calibrator integrity. Recalibrate with fresh calibrator if available.
High control high, low control normal 1-2s on high control Specific to high analyte concentration; possible non-linearity issue Check instrument linearity. Run backup analyzer if available. Verify no instrument saturation at high levels.
Results within control limits but consistently biased None (false pass) Systematic bias masked by wide control limits or insufficient controls Add third control level (middle concentration). Perform accuracy verification. Review control ranges.
Increased CV% (scatter) over past 2 weeks R-4s, larger ranges on repeat runs Operator fatigue, technique drift, or environmental instability Audit operator technique. Check environmental conditions. Rotate operators. Perform preventive maintenance.
QC fails but re-run passes immediately 1-2s or 1-3s (isolated) Random error or transient instrument issue Document as transient. No corrective action needed if QC confirms acceptable. Monitor for pattern recurrence.
QC fails after instrument maintenance 2-2s (all levels), shift pattern Maintenance effect on calibration or optical alignment Perform recalibration. Run full verification (low, mid, high controls). Compare pre/post-maintenance control values.
Gradual downward trend over 5 days 4-1s trending down Reagent lot deterioration or control material decay Replace control lot first. If trend continues, replace reagent. Check storage conditions. Verify expiration dates.
QC fails in morning only 1-2s, 1-3s (time-dependent) Instrument not warmed up adequately or environmental temperature variation Extend warm-up time in morning. Verify lab temperature overnight (check HVAC schedule). Allow instrument thermal stabilization.
One level fails, other level passes 1-3s or 1-2s on single level Random error or concentration-specific issue Re-run both levels. If single level persists, run fresh control. Check for instrument non-linearity at that concentration.
All analytes fail on same sample Multiple 2-2s or 1-3s (systematic) Instrument-wide issue or reagent contamination Run fresh control. Check reagent lot. Inspect instrument for contamination. Run instrument self-test.
QC passes but patient result seems wrong No QC violation (control bias issue) Control material not representative of patient sample; matrix effects Perform external quality assessment. Compare with reference lab. Review patient sample type for interferences.
Shift coincides with new operator starting 2-2s, persistent bias Operator technique error or change in procedure Observe operator technique. Provide hands-on training. Perform side-by-side comparison with experienced operator.
Spike after reagent cartridge changed 2-2s, 1-3s (all levels) New cartridge calibration or reagent batch variability Run quality verification if available. Check cartridge lot number. Recalibrate if required by instrument procedure.
Cluster of failures on Monday Multiple 1-2s or 1-3s on first runs Control material stability after weekend or instrument cool-down Extend warm-up Monday morning. Run control at start of day before patient samples. Verify storage conditions over weekend.
Erratic high/low pattern alternating R-4s, large range values Pipette accuracy issue or control mixing inconsistency Verify pipette calibration. Check probe/tip condition. Ensure adequate control mixing before aspiration.
QC high after power interruption 2-2s, sustained high shift Instrument temperature instability or calibration loss Allow extended warm-up after power restoration. Perform recalibration. Run verification before patient testing.
Negative result where impossible (e.g., potassium) 1-3s, direction-dependent Calibration error or electrode malfunction (chemistry analyzers) Check electrode status. Recalibrate with known calibrator. If persistent, schedule service call.
QC acceptable but poor between-run precision No QC violation (control range issue) Control acceptance range too wide; masking true imprecision Tighten control limits (narrow range). Implement tighter CV% limits. Add additional control levels.
Failure at specific control range level only 1-2s on mid or high control primarily Instrument non-linearity or concentration-dependent accuracy issue Run multiple controls across range. Plot accuracy vs concentration. Adjust calibration curve if available.
Gradual upward creep with new control lot 4-1s (subtle upward trend) Control lot baseline difference or deterioration from manufacturing Compare assigned values between lots. Run both old and new lot in parallel over 3-5 days. Contact manufacturer if concern.
QC fails intermittently (2-3 times per week) Random 1-2s or 1-3s without pattern Environmental fluctuation (temperature, humidity) or operator technique variation Log time of failures and correlate with room conditions. Monitor HVAC schedules. Provide retraining on consistent technique.
Complete failure of one QC rule (e.g., always 1-2s violations) Persistent 1-2s or 1-3s Control acceptance range too tight relative to true instrument precision Revalidate control acceptance ranges. Consider widening limits if instrument precision known. Document if limits are appropriate.
QC fine but patient results inconsistent with external lab No QC violation (systematic bias) Control material not representative of patient matrix; instrument calibration bias Perform external quality assessment. Run split samples with reference lab. Review calibrator traceability.
Sudden failure right after reagent delivered 2-2s, 1-3s (all levels, same direction) New reagent lot quality issue or improper shipment/storage Contact vendor; check reagent condition on arrival. Run quality verification. Request alternative lot if available.
QC low on cold mornings 1-2s, low control low in winter Lab temperature not adequate during overnight cool-down Adjust thermostat overnight. Extend warm-up time. Verify HVAC maintains minimum 18°C overnight.
Failure pattern matches sample run duration Increasing 1-2s violations through run Reagent or control degradation over time; temperature increase during run Check reagent stability in instrument at running temperature. Verify cooling/heating blocks function. Limit run time if needed.

Documentation Templates

Copy these templates to document QC failures and corrective actions in your laboratory.

QC Failure Investigation Report

QC FAILURE INVESTIGATION REPORT Laboratory: _________________________ Date: _________ Time: _________ Analyzer/Test: _________________________ QC Control Level: ___________ QC FAILURE DETAILS: Rule Violated: [ ] 1-2s [ ] 1-3s [ ] 2-2s [ ] R-4s [ ] 4-1s [ ] 10x [ ] Other: _____ Control Result: _________ μmol/L (or units) Expected Range: _________ - _________ μmol/L Control Mean: _________ SD: _________ Z-score: _________ Single or Multiple Levels Affected: [ ] Single [ ] Multiple [ ] All CONTROL MATERIAL VERIFICATION: Control Lot #: _____________ Expiration Date: ____________ Status: [ ] OK [ ] Expired Reconstitution Date: ____________ Freezing Cycles: _________ Status: [ ] OK [ ] Excessive Storage Temperature: _________ Storage Condition: [ ] OK [ ] Questionable Action Taken: [ ] Continue with same lot [ ] Replace with fresh lot REAGENT VERIFICATION: Reagent Lot #: _____________ Expiration Date: ____________ Status: [ ] OK [ ] Expired Recent Lot Change: [ ] No [ ] Yes (Date: _______) Reagent Condition: [ ] OK [ ] Discoloration [ ] Precipitate [ ] Other: _________ Reconstitution Date: ____________ Status: [ ] OK [ ] Excessive age Action Taken: [ ] Continue [ ] Replace reagent lot INSTRUMENT STATUS: Last Calibration: ____________ Status: [ ] OK [ ] Drift detected Last Maintenance: ____________ Type: _____________ Status: [ ] OK [ ] Post-service issue Error Logs Reviewed: [ ] No [ ] Yes - Issues found: ________________ Self-Test Result: [ ] Pass [ ] Fail - Issue: ________________________ Optical Check (if applicable): [ ] OK [ ] Contamination [ ] Misalignment ENVIRONMENTAL FACTORS: Laboratory Temperature: _________ °C (Target: 20-25°C) Status: [ ] OK [ ] Out of range Humidity: _________% RH (Target: 40-60%) Status: [ ] OK [ ] Out of range Ambient Conditions: [ ] Normal [ ] Drafts [ ] Direct sunlight [ ] HVAC issue Action Taken: ________________________________________________________ OPERATOR/TECHNIQUE REVIEW: Operator Performing Test: _____________ Experience Level: [ ] Experienced [ ] New [ ] Training needed Technique Review: [ ] Normal [ ] Deviation noted: ______________________ Recent Changes in Procedure: [ ] No [ ] Yes - Description: ________________ Training Recommendation: [ ] None needed [ ] Refresher recommended [ ] Hands-on retraining required CORRECTIVE ACTIONS TAKEN: [ ] Replaced control material - New lot #: ______________ [ ] Replaced reagent - New lot #: ______________ [ ] Performed recalibration [ ] Adjusted environmental conditions [ ] Cleaned/serviced instrument - Service performed: ____________ [ ] Retraining provided [ ] Modified standard operating procedure [ ] Other: _________________________________________________________________ VERIFICATION OF CORRECTION: Re-run QC Result: _________ μmol/L Status: [ ] Pass [ ] Fail Second Verification: _________ μmol/L Status: [ ] Pass [ ] Fail Patient Result Impact: [ ] Held pending resolution [ ] Released with comment [ ] Retested Patient Samples Affected: __________ (Time window: ____________) Clinical Significance: [ ] None [ ] Minor [ ] Significant ROOT CAUSE IDENTIFIED: [ ] Random error - Likely cause: ________________________________ [ ] Systematic error - Likely cause: ________________________________ [ ] Pre-analytical - Likely cause: ________________________________ [ ] Control/Reagent issue - Likely cause: ________________________________ [ ] Instrument malfunction - Likely cause: ________________________________ [ ] Operator technique - Likely cause: ________________________________ [ ] Environmental - Likely cause: ________________________________ PREVENTIVE ACTIONS (if applicable): [ ] Increased QC frequency [ ] Enhanced operator training [ ] Modified QC acceptance criteria [ ] Replacement of equipment component [ ] Procedure modification: __________________________________________ [ ] Other: __________________________________________________________ DOCUMENT REVIEW & APPROVAL: Investigator: _________________________ Signature: _________ Date: _______ Supervisor Review: _________________________ Signature: _________ Date: _______ Director Approval: _________________________ Signature: _________ Date: _______ FOLLOWUP (if needed): Followup Date: ___________ Outcome: ___________________________________

Corrective Action Record

CORRECTIVE ACTION RECORD Issue ID #: _____________ Date Identified: _________ Date Documented: _________ Product/Service/Process: _________________________________________________ Affected Department: _________________ Analyzer/Test: __________________ PROBLEM DESCRIPTION: QC Rule Violated: _________________ Magnitude of Deviation: ____________ Frequency: [ ] One occurrence [ ] Recurring (______ times) [ ] Systematic Patient Impact: [ ] None [ ] Potential [ ] Confirmed ROOT CAUSE ANALYSIS: Contributing Factor #1: __________________________________________________ Contributing Factor #2: __________________________________________________ Contributing Factor #3: __________________________________________________ Underlying Root Cause: ___________________________________________________ IMMEDIATE CORRECTIVE ACTIONS (Stop the Problem): Action #1: _________________________________ Completion Date: _________ Action #2: _________________________________ Completion Date: _________ Action #3: _________________________________ Completion Date: _________ Responsibility: _________________ Verify Effective: [ ] Yes [ ] No PERMANENT CORRECTIVE ACTIONS (Prevent Recurrence): Action #1: _________________________________ Target Completion: _________ Action #2: _________________________________ Target Completion: _________ Action #3: _________________________________ Target Completion: _________ Implementation Method: ___________________________________________________ PREVENTIVE ACTIONS (Future Prevention): Enhanced Monitoring: [ ] Increased QC frequency [ ] Additional controls Procedure Changes: [ ] Documented [ ] Training required Equipment: [ ] Scheduled maintenance [ ] Replacement recommended Training: [ ] Operator retraining [ ] Competency assessment EFFECTIVENESS VERIFICATION: Monitoring Period: From _________ To _________ QC Results Post-Correction: All within limits: [ ] Yes [ ] No Recurrence Observed: [ ] No [ ] Yes - Date: _________ Patient Results Post-Correction: Acceptable [ ] Yes [ ] No [ ] N/A External QA Results: Acceptable [ ] Yes [ ] No [ ] N/A APPROVAL & DOCUMENTATION: Initiator: _________________________ Signature: _________ Date: _________ Department Manager: _________________________ Signature: _________ Date: _________ Quality Manager: _________________________ Signature: _________ Date: _________ ANNUAL REVIEW: Last Reviewed: _________ Action Still Effective: [ ] Yes [ ] No Recommended Changes: _____________________________________________________

Patient Result Review Form

PATIENT RESULT REVIEW FORM (For Use When QC Was Out of Specification) QC Failure Date: _________ Test/Analyzer: _________________ Control Level: _____ QC Rule Violated: _____________ Control Result: _________ (Expected: _______) PATIENT SAMPLES AFFECTED: Testing Time Window: From _________ To _________ Number of Patient Samples: _________ Analyte(s): ________________________ IMPACT ASSESSMENT: Magnitude of Control Deviation: ___________% (High/Medium/Low) Direction: [ ] High [ ] Low [ ] Random Expected Impact on Patient Results: [ ] Minimal [ ] Moderate [ ] Significant Clinical Significance: [ ] Results still clinically valid [ ] Results questionable [ ] Results invalid DECISION MATRIX: Root Cause Identified: [ ] Yes [ ] No Corrective Action Completed: [ ] Yes [ ] No Verification of Correction: [ ] Pass [ ] Fail [ ] In progress DISPOSITION DECISION: [ ] RELEASE RESULTS - Reason: ______________________________________ - All samples released as reported - Clinical significance assessment: Results unchanged by deviation magnitude - Documentation: Results reported with comment (if applicable) [ ] RELEASE WITH COMMENT - Reason: ____________________________________ - Samples released with comment in LIS: "QC deviation on [date]; verify clinical correlation" - Physician notification recommended: [ ] Yes [ ] No - Specific analytes affected: _____________________________________ [ ] HOLD RESULTS - Reason: ________________________________________ - Samples held pending re-analysis or resolution - Expected resolution date: _________ - Alternative testing: [ ] Backup analyzer [ ] Reference lab [ ] Re-test [ ] RETESTING REQUIRED - Reason: ____________________________________ - Samples requiring re-analysis with acceptable QC - Retest date: _________ Results: __________________________________ - Final disposition: [ ] Released [ ] Still questionable PHYSICIAN NOTIFICATION: Notification Required: [ ] No [ ] Yes Physician: _________________________ Contact Date: _________ Method: __________ Message: "QC deviation identified on [date] for [test]. Your patient result [sample ID] is [valid/requires verification/being reanalyzed]. Please [action]." PATIENT SAFETY CONSIDERATIONS: Result Bias Direction: [ ] Toward higher risk [ ] Toward lower risk [ ] Neutral Critical Result Status: [ ] Not critical [ ] Critical - Notification: ____________ Repeat Testing Urgency: [ ] Routine [ ] Priority Clinical Outcome Risk: [ ] Low [ ] Moderate [ ] High - Escalate to: ____________ DOCUMENTATION: Results Released On: _________ Results Held On: _________ Superseded By: _________ (If retested) Record Filed: [ ] Patient chart [ ] Lab records [ ] Incident report REVIEWER SIGNATURE & APPROVAL: Reviewed By: _________________________ Signature: _________ Date: _________ (Authorized Laboratory Professional) Approved By: _________________________ Signature: _________ Date: _________ (Laboratory Director or Delegate)

Root Cause Analysis Summary

ROOT CAUSE ANALYSIS SUMMARY Date: _________ QC Issue ID: _________ Analyzer: __________________ EXECUTIVE SUMMARY (1-2 sentences): A QC [rule type] failure on [date] resulted from [root cause]. Corrective action [action taken] has resolved the issue. All affected patient results have been [disposition]. PROBLEM DEFINITION: What Happened: ________________________________________________________ When It Happened: _________ Where It Happened: __________________________ Scope: __________ Patient samples affected over __________ time period Severity: [ ] Low (isolated incident) [ ] Medium (recurring) [ ] High (widespread impact) INVESTIGATION PROCESS: Data Reviewed: [ ] QC trends [ ] Patient results [ ] Instrument logs [ ] Reagent lot records [ ] Environmental data Tools Used: [ ] QC rule analysis [ ] Control charting [ ] Fishbone diagram [ ] 5 Whys [ ] Other: _______ Interviews Conducted: [ ] Operator(s) [ ] Maintenance [ ] Management [ ] Supplier FISHBONE (5 CATEGORIES): People/Operator: _______________________________________________________ Procedure/Method: ______________________________________________________ Plant/Equipment: _______________________________________________________ Parts/Materials: _______________________________________________________ Environment: __________________________________________________________ WHY ANALYSIS (Root Cause): Why did QC fail? ________________________________________________________ Why was that a problem? __________________________________________________ Why did that condition exist? _____________________________________________ Why wasn't it caught earlier? ______________________________________________ ROOT CAUSE: ____________________________________________________________ EVIDENCE SUPPORTING ROOT CAUSE: Evidence #1: ____________________________________________________________ Evidence #2: ____________________________________________________________ Evidence #3: ____________________________________________________________ CORRECTIVE ACTIONS IMPLEMENTED: Immediate: ____________________________________________________________ Long-term: ____________________________________________________________ Preventive: ____________________________________________________________ EFFECTIVENESS CHECK (30-60 days post-action): QC Performance: [ ] Normal [ ] Improved [ ] Worsened [ ] No change Patient Result Quality: [ ] Acceptable [ ] Improved Recurrence: [ ] None [ ] Occasional [ ] Frequent Effectiveness: [ ] Effective [ ] Partially effective [ ] Ineffective - Escalate LESSONS LEARNED: What went well: ________________________________________________________ What could improve: ____________________________________________________ Procedure modifications made: ___________________________________________ Training reinforced: ____________________________________________________ APPROVED BY: QA Manager: _________________________ Signature: _________ Date: _________ Laboratory Director: _________________________ Signature: _________ Date: _________

QC Troubleshooting and ISO 15189:2022

Clinical laboratory standards require documented investigation and corrective action for all QC failures.

Relevant ISO 15189:2022 Requirements

Clause 7.2: Management of pre-analytical and analytical phases

7.2.1 QC: "The laboratory shall establish and implement procedures to ensure the quality of analytical results including procedures for pre- and post-analytical phases."

  • Define QC procedures, acceptance criteria, and corrective actions
  • Document all QC results (passes and failures)
  • Implement immediate corrective actions for QC failures
  • Assess impact on patient results

Clause 8.4: Nonconforming analytical results

8.4.1 Nonconforming Results: "The laboratory shall identify and manage analytical results that do not meet its defined criteria for acceptability."

  • Identify QC failures promptly
  • Document the nonconforming situation
  • Assess the extent of the problem (how many patient samples affected)
  • Determine corrective action
  • Record and communicate with relevant parties

Clause 8.5: Corrective action

8.5.1 General: "The laboratory shall establish and implement procedures to take corrective action for nonconformities identified, including QC failures."

  • Root cause analysis methodology (5 Whys, fishbone, etc.)
  • Documented corrective action plan
  • Implementation timeline and responsibility assignment
  • Verification that corrective action was effective
  • Prevention of recurrence (preventive action)

Clause 8.6: Management review

8.6.2 Input to Management Review: "Quality management system performance shall be evaluated, including... nonconformities and corrective actions taken."

  • Regular review of QC failures (monthly or quarterly)
  • Trend analysis across multiple failures
  • Effectiveness of corrective actions
  • Identification of systemic issues requiring wider intervention

Demonstrating Compliance to Assessors

During Laboratory Audits (NABL, CAP, CLIA)

  • Show QC Records: Maintain 1-2 years of QC results demonstrating control and detection of failures
  • Show Failure Documentation: Provide examples of properly investigated QC failures with corrective action records
  • Demonstrate Root Cause Analysis: Assessors will ask: "What was the root cause?" Have documented evidence
  • Show Patient Safety Measures: Document how you assessed and managed impact on patient results
  • Show Effectiveness Verification: Document that corrective actions work (QC improved post-correction)
  • Show Training Records: Training documentation for operators involved in failures and retraining provided

Key Questions Assessors Ask:

  • "How often do you have QC failures?" (Expected answer: rare, investigated thoroughly)
  • "How do you investigate QC failures?" (Expected: systematic, documented process)
  • "Can you show me a recent failure investigation?" (Be prepared to present complete documentation)
  • "How do you determine if patient results are affected?" (Show clear decision-making process)
  • "What corrective actions did you take?" (Show documented implementation and verification)
  • "How do you prevent failures from happening again?" (Show preventive actions and trend monitoring)

Documentation Best Practices for Compliance

  • Complete: Every QC failure must have a complete investigation record
  • Timely: Investigation should be documented within 24-48 hours of failure
  • Objective: Record facts (control values, environmental conditions) not opinions
  • Signed: Investigation must be signed by investigator and approved by supervisor/director
  • Traceable: Link QC failure to specific lot numbers, operators, and patient samples
  • Evidence-based: Root cause should be supported by data/investigation findings
  • Actionable: Corrective actions must be specific, assigned, and measurable
  • Verified: Document that corrective action resolved the problem

NABL- and CAP-Specific Considerations

NABL (National Accreditation Board for Testing and Calibration Laboratories)

NABL follows ISO 15189 closely. Additional requirements:

  • Internal audits must evaluate QC failure trends
  • Corrective actions must be tracked to closure
  • Staff competency must be maintained (training records for all personnel)
  • Supplier controls (reagents, controls) must be verified

CAP (College of American Pathologists)

CAP inspection checklist includes:

  • QC rules defined and appropriate for analytes
  • QC failures documented with dates and results
  • Corrective action plan documented and implemented
  • Patient result impact assessment and communication
  • Preventive actions to avoid recurrence
  • Regular review of QC performance (Monthly QC Summary recommended)

Automate QC Monitoring and Documentation

Never miss a QC violation. QCnomics provides real-time QC monitoring, automatic rule detection, and built-in templates for investigation and corrective action documentation—all integrated with your lab's existing workflow.