Home / Automated Westgard multi-rule evaluation software
CLSI C24-A4 · Real-time IQC

Automated Westgard multi-rule evaluation software

Stop hand-checking QC against Westgard rules. QCnomics evaluates every control result the moment it is entered — within-run, across-run and across-material — and flags true violations while suppressing false rejections.

Try QCnomics

What are Westgard multi-rules?

Westgard multi-rules are a set of statistical control rules used to decide whether an analytical run is in control. Instead of a single 2SD limit (which rejects too often), a combination of rules improves error detection while limiting false rejection. The common set includes 12s (warning), 13s, 22s, R4s, 41s, 8x and 10x.

Within-run, across-run and across-material

Rules are not all applied the same way. R4s is a within-run rule (range across levels in the same run). 22s, 41s, 10x can be applied across consecutive runs and, where appropriate, across materials (pooling control levels using SDI). Applying rules at the wrong scope causes both missed errors and false alarms — QCnomics applies each rule at its correct scope automatically.

CLSI C24-A4 phased QC

QCnomics implements the CLSI C24-A4 phased approach: a verification phase for the first runs of a new lot, a provisional phase as data accumulates, and a fixed phase once the mean and SD are locked (typically after 20 in-control runs). The active rule set and limits are selected automatically based on run count and lot.

What QCnomics does for you

Frequently asked questions

Which Westgard rules should a clinical chemistry lab use?

A common, well-balanced set is 1-3s / 2-2s / R-4s / 4-1s / 10x, with 1-2s as a warning only. QCnomics lets you enable the rules appropriate to each analyte and applies them at the correct within-run/across-run scope.

Does QCnomics reduce false rejections?

Yes. By using multi-rule logic instead of a blanket 2SD cutoff and by applying rules at the correct scope and phase, it flags genuine out-of-control situations while avoiding unnecessary repeat testing.

Is it aligned with CLSI C24-A4?

Yes — the verification, provisional and fixed phases and the mean/SD locking follow CLSI C24-A4 guidance.

Do this automatically in QCnomics

Clinical-grade internal quality control and ISO 15189:2022 compliance in one platform.